Product Nee: PM400F
PM400F Flashlight Photometer is an instrument designed specifically for the instantaneous photometric testing of pulsed and flashing light sources. It can operate as a standalone unit or in conjunction with LISUN’s LSG series goniophotometers for integrated testing. It is suitable for measuring the instantaneous photometric characteristics and effective luminous intensity of various pulsed light sources—including aviation obstruction lights, aircraft anti-collision lights, navigation lights, traffic warning lights, special vehicle warning lights, pulsed LEDs, flares, signal flares, and pyrotechnics. The device supports type testing and factory acceptance testing for flashing warning lights used in aviation, marine, automotive, and other civil applications. Featuring a detector with excellent V(λ) spectral response matching, it is compatible with diverse flash waveforms—such as square waves, irregular pulses, and pulse trains—and meets the metrological requirements of relevant domestic and international standards for flashing light sources. sources.
Normen:
| Normen nr. | Normen Naam |
| ECE R65 | Uniforme bepalingen betreffende de goedkeuring van speciale waarschuwingslichten voor motorvoertuigen |
| SAE J595 | Directional Flashing Optical Warning Devices for Authorized Emergency, Maintenance, and Service Vehicles |
| SAE J845 | Optical Warning Devices for Authorized Emergency, Maintenance, and Service Vehicles |
| MH/T 6012‑2015 | 航空障碍灯 |
| HB 6490‑1991 | 飞机航行灯和防撞灯通用规范 |
| JT/T 761‑2022 | 航标灯通用技术条件 |
| JJF 1330‑2011 | 瞬态有效光强测定仪校准规范 |
Specificaties:
| Meetbare parameters | Flash frequency, flash duration, effective luminous intensity, instantaneous maximum luminous intensity, instantaneous minimum luminous intensity, average luminous intensity, etc. Please click here for a sample report. |
| A / D-omzetting | 16-bit high-speed A/D with a maximum sampling rate of 1 MSPS and support for storing up to 20 million sample points. |
| Bereik van verlichtingssterktemeting | 0.0001 lx to 200 klx, automatic switching between 4 ranges |
| Photometric Probe Nauwkeurigheid |
CIE Class A (Class L Optional) |
| Testing Time Window | 100 ms~10 s |
| Maximum Temporal Resolution | 10 μs |
| Communicatie Interface | USB, can be used in conjunction with the LISUN Goniofotometers uit de LSG-serie. |
| Display and Power Supply | LCD display screen, DC 5V / 1A |
| Afmetingen en gewicht | L*W*H: 170*120*120 mm、around 1 kg |